Environmental Requirements

Room Temperature 15º- 25ºC
Humidity 20% to 85%
Heat Dissipation 773W (2637 BTU/Hr) during steady operation

 

Electrical Requirements

Analyzer 200-240 VAC, 50-60 Hz, 30A (dedicated circuit required)
RPI US 120V AC, 60 Hz, 650 Watts

 

Other

Sample Tube Bar Code Specifications Code 39, Code 128 (ISBT 128), Codabar
Sample Size Tubes 12 x 75 mm, 13 x 75 mm, 13 x 100 mm, 16 x 75 mm, 16 x 100 mm
Operating System Windows® XP
Water Requirement Type I Reagent-Grade Water
Waste Collection No special plumbing required; self-contained

 