Specifications
Requirements
Environmental Requirements
| Room Temperature | 15º- 25ºC |
| Humidity | 20% to 85% |
| Heat Dissipation | 773W (2637 BTU/Hr) during steady operation |
Electrical Requirements
| Analyzer | 200-240 VAC, 50-60 Hz, 30A (dedicated circuit required) |
| RPI | US 120V AC, 60 Hz, 650 Watts |
Other
| Sample Tube Bar Code Specifications | Code 39, Code 128 (ISBT 128), Codabar |
| Sample Size Tubes | 12 x 75 mm, 13 x 75 mm, 13 x 100 mm, 16 x 75 mm, 16 x 100 mm |
| Operating System | Windows® XP |
| Water Requirement | Type I Reagent-Grade Water |
| Waste Collection | No special plumbing required; self-contained |